FDA recall Z-2570-2025

Philips Medical Systems (Cleveland) Inc · Class II · device

Product

Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacle TumorLOC 16.2 Model 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Pinnacle TumorLOC 16.2.1 Model 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Pinnacle TumorLOC 16.0.2 Model 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Pinnacle TumorLOC 14.0 Model 870200 UDI code: N/A Pinnacle3 Radiation Therapy Planning System is a software package intended to provide planning support for the treatment of disease processes. The full Pinnacle3 Radiation Therapy Planning System software package provides planning support for the treatment of disease processes, utilizing photon, electron and brachytherapy techniques.

Reason for recall

Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.

Distribution

U.S Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.

Key facts

Status
Ongoing
Initiation date
2025-08-05
Report date
2025-09-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Gainesville, FL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2570-2025