# FDA recall Z-2573-2023

> **Landauer** · Class II · device recall initiated 2023-07-12.

## Product

NanoDots, Model Numbers:    a) BC30023, description:  QC Set For Microstar Dots;   b) BC30083, description:  nanoDot QC Set - 80kVp - Sale;   c) BC30084, description:  nanoDot QC Set - 80kVp - No Charge;   d) BC30088, description:  nanoDot Calibration Set - 80kVp - No Charge;   e) BC30095, description:  Dot, D1DNN, in carrier sealed in bag;   f) BC30110, description:  Calibration set, nanoDot;   g) BC30141, description:  nanoDot QC Set - Cs137 - Sale;   h) BC30142, description:  nanoDot Calibration set - Cs137 - Sale; radiation monitoring dosimeter used with the microSTAR readers

## Reason for recall

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

## Distribution

Worldwide

## Key facts

- **Recall number:** Z-2573-2023
- **Recalling firm:** Landauer
- **Classification:** Class II
- **Product type:** device
- **Status:** Completed
- **Initiation date:** 2023-07-12
- **Report date:** 2023-09-20

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Glenwood, IL, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2573-2023

## Citation

> AI Analytics. FDA recall Z-2573-2023. Retrieved 2026-06-24 from https://api.ai-analytics.org/recall/Z-2573-2023. Source: US FDA. Licensed CC0.

---