FDA recall Z-2575-2018

Randox Laboratories Ltd. · Class II · device

Product

Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems.

Reason for recall

The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.

Distribution

Worldwide Distribution - US in the state of WV. Foreign distribution to Australia, Chile, China, Germany, Hungary, Ireland, Italy, Moldova, Netherlands, Peru, Poland, Saudi Arabia, Slovakia, Spain, Sultanate of Oman, UAE, UK, and Vietnam

Key facts

Status
Terminated
Initiation date
2018-06-08
Report date
2018-08-08
Termination date
2018-11-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Crumlin (North), N/A, Ireland