FDA recall Z-2575-2021

Medline Industries Inc · Class II · device

Product

NAMIC Stopcocks: (1) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015002, UPN H965700150021, Three-way Stopcock with Rotating Collar, Rx only, Sterile; (2) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015003, UPN H965700150031, Three-Way Stopcock with Fixed Male Luer Lock, Rx only, Sterile; (3) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015012, UPN H965700150121, Four-way Stopcock with Rotating Collar, Rx only, Sterile; (4) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015013, UPN H965700150131, Four-way Stopcock with Fixed Male Luer Lock, Rx only, Sterile; (5) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015015, UPN H965700150151, One-way Stopcock with Collar, Rx only, Sterile; (6) NAMIC STOPCOCK, 400 psi (28 bar), REF 70035002, UPN H965700350021, Three-way Stopcock with Rotating Collar, Rx only, Sterile; (7) NAMIC STOPCOCK, 400 psi (28 bar), REF 70035003, UPN H965700350031, Three-way Stopcock with Fixed Male Luer Lock, Rx only, Sterile; (8) NAMIC STOPCOCK, 400 psi (28 bar), REF 7003

Reason for recall

Potential for sterile barrier breach. Microscopic pinholes within the sterile pouch could potentially lead to non-sterile conditions.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Canada, Australia, Japan, Latvia, Lebanon, Malaysia, Netherlands, Pakistan, People's Republic of China, Panama, Singapore, South Africa, South Korea, Thailand, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2021-08-06
Report date
2021-10-06
Termination date
2025-08-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2575-2021