FDA recall Z-2575-2023

Landauer · Class II · device

Product

Reader, nanoDot adapter, EU Model Number VINLADA003; radiation monitoring dosimeter used with the microSTAR readers

Reason for recall

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Distribution

Worldwide

Key facts

Status
Completed
Initiation date
2023-07-12
Report date
2023-09-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Glenwood, IL, United States

Primary source