FDA recall Z-2579-2018

GE Healthcare, LLC · Class II · device

Product

Discovery MI Digital Ready

Reason for recall

DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Distribution

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Key facts

Status
Terminated
Initiation date
2018-05-24
Report date
2018-08-08
Termination date
2019-02-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2579-2018