# FDA recall Z-2586-2024

> **EYE COMFORT CARE LLC** · Class II · device recall initiated 2024-02-08.

## Product

Brand Name: UNclog Eye Mask  Product Name: UNclog Eye Mask  Model/Catalog Number: AG23JW24 Serial # ag22  Product Description: Heated eye mask  Component: Heating unit, wire

## Reason for recall

Due to a change in the safety clips for the heated eye masks, the wires to the safety clips may detach as a result of tugging and pulling and may cause a short in the heating element of the mask and could potentially cause burn injuries.

## Distribution

U.S. Nationwide distribution in the states of AL, AZ, CA, CT, FL, KS, KY, MA, ME, MT, NC, ND, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, and WI.

## Key facts

- **Recall number:** Z-2586-2024
- **Recalling firm:** EYE COMFORT CARE LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Completed
- **Initiation date:** 2024-02-08
- **Report date:** 2024-08-21

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Gales Creek, OR, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2586-2024

## Citation

> AI Analytics. FDA recall Z-2586-2024. Retrieved 2026-06-10 from https://api.ai-analytics.org/recall/Z-2586-2024. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*