FDA recall Z-2603-2018

Medtronic Sofamor Danek USA Inc · Class II · device

Product

Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.MULTI-AXIAL SCREW FOR 5.5/6.0MM RODS, REF 55840005045

Reason for recall

This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly indicate a 5.5 mm diameter screw when the actual screw diameter aligns with 5.0 diameter screw size requirements.

Distribution

US

Key facts

Status
Terminated
Initiation date
2018-07-09
Report date
2018-08-15
Termination date
2021-05-07
Voluntary/Mandated
Voluntary: Firm initiated
Location
Memphis, TN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2603-2018