FDA recall Z-2611-2025

Medline Industries, LP · Class I · device

Product

Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbers BD710DF282CRH BD710DF282RRH BD710FJ282CRH BD710FJ282RRH

Reason for recall

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Distribution

Domestic US distribution nationwide. International distribution pending. No international distribution.

Key facts

Status
Ongoing
Initiation date
2025-07-31
Report date
2025-10-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2611-2025