# FDA recall Z-2615-2023

> **Abbott Medical** · Class II · device recall initiated 2023-08-16.

## Product

Gallant HF Cardiac Resynchronization Therapy Defibrillator:  HF CRT-D REF CDHFA500Q    Neutrino HF Cardiac Resynchronization Therapy Defibrillator:  NxT HF CRT-D REF CDHFA600Q    Entrant HF Cardiac Resynchronization Therapy Defibrillator :  HF CRT-D REF CDHFA300Q    The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dual chamber (also known as DR if rate responsive) device that is connected to a right atrial (RA) pacing lead and right ventricular defibrillation lead or single chamber ICD (also known as VR if rate responsive) with only a defibrillation lead in the right ventricle (RV). The ICD, more generally known as an implantable pulse generator (IPG), is typically implanted in a subcutaneous device pocket located in the pectoral region and is attached to the pacing/defibrillator leads that are implanted thr

## Reason for recall

Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication.  In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

## Distribution

U.S.: Nation-wide  O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Nepal, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom.  Latin America - Pending

## Key facts

- **Recall number:** Z-2615-2023
- **Recalling firm:** Abbott Medical
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-08-16
- **Report date:** 2023-09-27

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Sylmar, CA, United States