FDA recall Z-2625-2023

Remote Diagnostic Technologies Ltd. · Class II · device

Product

Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF 01-2044

Reason for recall

Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia.

Distribution

US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME, MI, MN, MO, MS, MT, NC , ND, NE, NJ, NM , NV, NY, OH, OK, OR, PA, TN, TX, UT, VA , WA, WI, WV, WY. OUS: AE, AT, AU, BE, BN, BO, CA, CH, CZ, DE, DK, ES, FR, GB, HK, IE, IL, IS, IT, KY, LU, MY, NL, NO, OM, PT, RO, SA, SE, SG, SI, TH, YT, ZA

Key facts

Status
Ongoing
Initiation date
2023-08-24
Report date
2023-09-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Farnborough, United Kingdom

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2625-2023