FDA recall Z-2629-2023

WOM World Of Medicine AG · Class II · device

Product

Aquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopies and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus. Models AQL-100PBS and AQL-100P

Reason for recall

The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached and may result in fluid overload

Distribution

Nationwide Foreign: Canada, Australia, New Zealand, Belgium, France, Germany, Ireland, Netherlands, Spain, Sweden, Norway, Switzerland, UK, Kuwait, Qatar, Turkey, UAE

Key facts

Status
Ongoing
Initiation date
2023-07-17
Report date
2023-10-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Ludwigsstadt, N/A, Germany

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2629-2023