FDA recall Z-2633-2017

Elekta, Inc. · Class II · device

Product

MOSAIQ(R) Oncology Information System, Versions 2.50.05 and higher Product Usage: MOSAIQ¿ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed. Users can configure MOSAIQ¿ for Medical Oncology use, Radiation Oncology use, or the two together. It lets users:  Assemble electronic patient charts and treatment plans, order diagnostic tests, and prescribe medications.  Generate and keep medication formulary lists and calculate applicable medication dosages for medical oncology.  Import, view, annotate, adjust, enhance, manage and archive images.  Compare radiation treatment plans and evaluate dose coverage.  Design leaf plans for operation with radiotherapy treatment machines that have multileaf collimators.  Make sure radiation treatment plans imported from treatment planning systems agree with treatme

Reason for recall

Retention of Outdated Information in an Order Set. In the Care Plan Library, double clicking to add or delete a Component/Procedure to an Order Set does not trigger the update of the associated Care Plans.

Distribution

Worldwide Distribution - US Nationwide and the countries of ANTIGUA AND BARBUDA, AUSTRALIA, AUSTRIA, BAHRAIN, CANADA, CHEZ REPUBLIC, GERMANY, HUNGARY, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWEDEN, AND UK

Key facts

Status
Terminated
Initiation date
2017-06-21
Report date
2017-07-05
Termination date
2018-07-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Atlanta, GA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2633-2017