FDA recall Z-2640-2023

3M Company - Health Care Business · Class II · device

Product

3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes

Reason for recall

Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.

Key facts

Status
Ongoing
Initiation date
2023-08-22
Report date
2023-10-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Paul, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2640-2023