FDA recall Z-2659-2023

Olympus Corporation of the Americas · Class II · device

Product

Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN Thoracic Navigation System SYS-3230 ig4 System, 230V, 50Hz / SPiN Thoracic Navigation System 0095 SYS-3000 SPiN Thoracic Navigation System SYS-2400 SPiN Drive System SYS-2230 ig4 System V2, 230V, 50Hz SYS-2000 ig4 Image Guides System SYS-1500 SPiN View" System SYS-1000 ig4 Image Guides System SYS-0220 ig4 System, 220V, 50Hz SYS-0185 SPiN Planning Laptop Workstation SYS-0128 Planning Workstation w/Monitor SYS-0002 ig4 System INS-7150 SPiN Vision Scope Adapter

Reason for recall

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

Distribution

Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.

Key facts

Status
Ongoing
Initiation date
2023-08-10
Report date
2023-10-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Center Valley, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2659-2023