FDA recall Z-2669-2020

Becton Dickinson & Company · Class II · device

Product

SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI 00801741066214), LH-0032YN (UDI 00801741066238), LH-0033YN (UDI 00801741066252), LH-0034YN (UDI 00801741066276), LH-0035YN (UDI 00801741066290 ), LH-0036YN (UDI 00801741066313), LH-0037YN (UDI 00801741066337), LH-0038YN (UDI 00801741066351), LH-0039YN (UDI 00801741066313), LH-0040YN (UDI 00801741066399)

Reason for recall

Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

Distribution

Worldwide Distribution

Key facts

Status
Terminated
Initiation date
2020-06-19
Report date
2020-07-29
Termination date
2022-07-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Franklin Lakes, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2669-2020