FDA recall Z-2686-2020

Mevion Medical Systems, Inc. · Class II · device

Product

MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients

Reason for recall

The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.

Distribution

MO, NJ, OK, FL, OH, DC, Netherlands

Key facts

Status
Terminated
Initiation date
2020-06-23
Report date
2020-07-29
Termination date
2020-12-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Littleton, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2686-2020