FDA recall Z-2688-2017

Siemens Medical Solutions USA, Inc · Class II · device

Product

Sensis Vibe Systems with Software Version VD10B, Model Numbers 10765502, 10910620, 11007641, 6648161 --- Programmable diagnostic computer,

Reason for recall

Software error. In Sensis Vibe systems with software version VD10B, a software error can result in: problems generating a report and/ or - information from different examinations of the same patient being combined into one report. --- The error causes information from two examinations to be combined into one report. This does not result in a hazardous situation; however, if the erroneous report has been used for treatment or diagnosis, this can potentially result in an incorrect treatment or diagnosis.

Distribution

US Distribution to the states of : IN, MI, IA

Key facts

Status
Terminated
Initiation date
2017-05-16
Report date
2017-07-12
Termination date
2018-08-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2688-2017