FDA recall Z-2702-2017

Penumbra Inc. · Class I · device

Product

Penumbra 3D Revascularization Device It is indicated for used in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral-M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Reason for recall

Penumbra has identified an issue in these four lots involving a raw material component of the delivery wire. This issue could result in breakage of the delivery wire, which could potentially lead to serious patient injury or death.

Distribution

Nationwide Distribution to the following states: AZ, CA, CO, FL, IN, MA, MI, MN, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WV

Key facts

Status
Terminated
Initiation date
2017-06-09
Report date
2017-07-26
Termination date
2019-05-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Alameda, CA, United States