FDA recall Z-2702-2020

Smith & Nephew, Inc. · Class II · device

Product

Smith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated for use in surgical arthroscopic rotator cuff repair procedures. REF/ Catalog Number : 014771

Reason for recall

The T-Fix sterile package contains only one sleeve instead of two, The device will not perform as intended for the second bone tunnel due to the absence of thesleeve

Distribution

Worldwide distribution - US Nationwide distribution including in the state of CO and the countries of AT, BE, CH, CL, DE, DK, ES, FR, GB, IT. Expanded 8/28/20 recall: 2nd Field Action: AU, CH, DE, FR, GB, IT, NL, PL, PT, & ZA

Key facts

Status
Ongoing
Initiation date
2020-06-24
Report date
2020-08-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Andover, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2702-2020