FDA recall Z-2707-2017

Merge Healthcare, Inc. · Class II · device

Product

The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.

Reason for recall

If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.

Distribution

Distribution US nationwide.

Key facts

Status
Terminated
Initiation date
2016-03-10
Report date
2017-07-19
Termination date
2020-03-06
Voluntary/Mandated
Voluntary: Firm initiated