FDA recall Z-2722-2020

CareFusion 303, Inc. · Class I · device

Product

Alaris SpO2 Module Model 8210 and Model 8220, a modular infusion pump and monitoring system

Reason for recall

Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.

Distribution

Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH

Key facts

Status
Terminated
Initiation date
2020-06-30
Report date
2020-08-12
Termination date
2024-04-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2722-2020