FDA recall Z-2722-2024

Stryker Corporation · Class II · device

Product

Stryker iBur 3.0mm Precision Round, Distal Bend- Intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-009-030

Reason for recall

Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.

Key facts

Status
Completed
Initiation date
2024-07-24
Report date
2024-09-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Portage, MI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2722-2024