FDA recall Z-2730-2017

Beckman Coulter Inc · Class II · device

Product

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.

Reason for recall

The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.

Distribution

US Nationwide Distribution in the states of AL, AR, AK, CA, CO, DE, FL, GA, ID, IL, IN, IO, KS, KY, LA, MA, MD, MS, MI, ME, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WV, WI, and Puerto Rico Brunei, Canada, China, Czech Republic, France, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Lebanon, Mexico, Netherlands, Panama, Paraguay, Philippines, Qatar, South Africa, Taiwan, Thailand, Turkey, United Arab Emirates, & United Kingdom.

Key facts

Status
Terminated
Initiation date
2017-05-15
Report date
2017-07-19
Termination date
2020-09-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brea, CA, United States

Primary source