FDA recall Z-2736-2020

CareFusion 303, Inc. · Class I · device

Product

Alaris Pump Module Model 8100

Reason for recall

Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame assembly. These may prevent devices from delivering accurate amounts of fluid, which may result in an over infusion, free-flow conditions, or under infusion.

Distribution

US: CA, FL, IN, TX, NC, LA, SC, VA, PA, OK, AZ, IL, KY, UT, AL, GA, TN, MA, MI, OH, AK, OR, MS, CO, NY, WI, KS, MO, MD, NV, AR, AS, ID, VT, NE, NJ, DC, HI, CT, MT, WV, NH, WA, DE, ME, NM, MN, WY, IA, SD, ND, MP, PR, GU. OUS: NL, MX, AU, MY, NZ, ZA, GB, FR, CH, CN, SG, TW, CA, IL, KW, QA, DE, BH, IN, SA, GI, AE, PK, PH, BE

Key facts

Status
Terminated
Initiation date
2020-06-30
Report date
2020-08-12
Termination date
2024-04-23