FDA recall Z-2741-2018

Compass Health Brands (Corporate Office) · Class I · device

Product

CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to an end user s face and a CPAP machine, to treat end users with sleep apnea. The replacement cushion is a non-durable component that requires replacement every 3 - 6 months. The cushion is used for both comfort and to provide a seal between the face and the mask.

Reason for recall

A May 2015 design change in the cushion seal replacement part and accompanying elbow replacement part resulted in incompatibility of the new cushion seal with the previous design of the elbow. Use of the new cushion seal with the previous design elbow could lead to unacceptable CO2 porting. As there was no model number change for this design change, customers may not be aware of the incompatibility.

Distribution

US nationwide distribution to AL, CT, FL, GA, KY, MS, NC, NY, TN, and TX.

Key facts

Status
Terminated
Initiation date
2018-06-18
Report date
2018-08-29
Termination date
2020-05-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Middleburg Heights, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2741-2018