FDA recall Z-2742-2020

OrthoPediatrics Corp · Class II · device

Product

Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease Model Number: 00-1300-6053

Reason for recall

Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended fracture of the device

Distribution

US Nationwide distribution including in the states of Al, CA, FL, GA, KY, MI, MO, NC, NY, OH, TX.

Key facts

Status
Ongoing
Initiation date
2020-07-06
Report date
2020-08-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States