# FDA recall Z-2747-2020

> **Siemens Healthcare Diagnostics, Inc.** · Class II · device recall initiated 2020-06-30.

## Product

DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 -  HC0015BZZWW external accessory to DCA Vantage Analyzer US

## Reason for recall

If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit.    Also,  if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.

## Distribution

Nationwide

## Key facts

- **Recall number:** Z-2747-2020
- **Recalling firm:** Siemens Healthcare Diagnostics, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2020-06-30
- **Report date:** 2020-08-12
- **Termination date:** 2020-11-17

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Newark, DE, United States