FDA recall Z-2748-2017

Becton Dickinson & Company · Class II · device

Product

BD SafetyGlide" Injection Needle Catalog #305917 The SafetyGlide " needle is intended to be used for general purpose injection, aspiration of fluids from parts of the body below the surface of the skin, injection into intravenous ports and tubings external to the patient. The needlestick prevention feature of the device contains a mechanism that covers the needlepoint after use. In the activated position the needle cover guards against accidental needle sticks during normal handling and disposal of the used needle/syringe combination

Reason for recall

BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.

Distribution

Nationwide Distribution

Key facts

Status
Terminated
Initiation date
2016-09-16
Report date
2017-07-26
Termination date
2018-03-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Franklin Lakes, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2748-2017