FDA recall Z-2751-2018

Terumo Cardiovascular Systems Corporation · Class II · device

Product

Monitor with one Blood Parameter Module and one HSat Probe; 500AHCT Monitor with two Blood Parameter modules and one HSat Probe; 500AVHCT Product Usage: The CDI System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. In addition, calculated values of base excess, bicarbonate, oxygen saturation, and oxygen consumption may also be provided. These parameters are displayed at either actual temperature or adjusted to 37¿C. For documentation purposes, the system 500s integral printer provides a hard copy of displayed parameters.

Reason for recall

H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.

Distribution

Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC, NM, NY, OH, SC, TN, TX, WA, and WV. Foreign distribution to Middle East, Latin America, Europe, Japan, Canada, and Mexico.

Key facts

Status
Terminated
Initiation date
2018-06-26
Report date
2018-08-22
Termination date
2020-04-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Ann Arbor, MI, United States

Primary source