FDA recall Z-2755-2017

Medtronic Navigation, Inc. · Class II · device

Product

Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling Catheter with 3M(TM) SteriStrip(TM), (1) Pump Tubing Set, (1) Drainage Bag. Medtronic Navigation, Inc. Louisville, CO 80027

Reason for recall

Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.

Distribution

US Distribution to the states of : CA, GA, MN, FL, NY, IL, TX

Key facts

Status
Terminated
Initiation date
2017-06-07
Report date
2017-08-02
Termination date
2018-06-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Louisville, CO, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2755-2017