FDA recall Z-2760-2026

Ferno-Washington Inc · Class II · device

Product

Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091

Reason for recall

Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, GA, IA, ID, IL, IN, KY, LA, MD, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Poland, Netherlands, Canada, Germany, Norway, United Kingdom, Denmark and New Zealand.

Key facts

Status
Ongoing
Initiation date
2026-06-19
Report date
2026-07-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wilmington, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2760-2026