# FDA recall Z-2771-2026

> **Boston Scientific Corporation** · Class II · device recall initiated 2026-06-19.

## Product

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

## Reason for recall

Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate",  the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.

## Distribution

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.

## Key facts

- **Recall number:** Z-2771-2026
- **Recalling firm:** Boston Scientific Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-06-19
- **Report date:** 2026-07-29

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Maple Grove, MN, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2771-2026

## Citation

> AI Analytics. FDA recall Z-2771-2026. Retrieved 2026-09-03 from https://api.ai-analytics.org/recall/Z-2771-2026. Source: US FDA. Licensed CC0.

---