FDA recall Z-2778-2026

Medtronic MiniMed, Inc. · Class II · device

Product

MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

Reason for recall

For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

Distribution

Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.

Key facts

Status
Ongoing
Initiation date
2026-06-16
Report date
2026-07-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northridge, CA, United States