FDA recall Z-2795-2018

Hamilton Medical AG · Class II · device

Product

Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.

Reason for recall

An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.

Distribution

US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.

Key facts

Status
Terminated
Initiation date
2018-03-09
Report date
2018-08-22
Termination date
2019-03-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bonaduz, Switzerland

Primary source