FDA recall Z-2803-2017

Quidel Corporation · Class II · device

Product

InflammaDry(R) AN RPS DIAGNOSTIC SOLUTION, Sample Collectors InflammaDry(R) is a rapid, immunoassay test for the visual, qualitative, in vitro detection of elevated levels of the MMP-9 protein in human tears, from patients suspected of having dry eye. InflammaDry is to be used to aid in the diagnosis of dry eye, in conjunction with other methods of clinical evaluation. This test is intended for prescription use at point-of-care sites.

Reason for recall

In-house testing revealed the presence of plastic spurs at the tips of some Sample Collectors.

Distribution

Nationwide Distribution including AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TX, VA, WA, & WV.

Key facts

Status
Terminated
Initiation date
2017-06-22
Report date
2017-08-09
Termination date
2019-06-07
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States