FDA recall Z-2807-2026

Medtronic Perfusion Systems · Class II · device

Product

EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77522

Reason for recall

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-06-24
Report date
2026-08-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brooklyn Park, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2807-2026