FDA recall Z-2816-2020

Synthes (USA) Products LLC · Class II · device

Product

DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula, particularly in osteopenic bone. Part Number 241.381

Reason for recall

May have one or more dimensional features out of tolerance includes: thread pitch diameter depth, thread minor diameter depth, plate thickness and end length out of tolerance that can potentially result in a reduced locking strength of an LCP screw

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2020-07-21
Report date
2020-08-26
Termination date
2022-02-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
West Chester, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2816-2020