FDA recall Z-2819-2020

Shanghai United Imaging Healthcare Co., Ltd. · Class II · device

Product

Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Reason for recall

In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.

Distribution

US Nationwide distribution including in the state of TX.

Key facts

Status
Terminated
Initiation date
2020-03-26
Report date
2020-08-26
Termination date
2023-08-02
Voluntary/Mandated
Voluntary: Firm initiated