FDA recall Z-2895-2020

Karl Storz Endoscopy · Class II · device

Product

Flexible DCI Intubation Fiberscope, Part: 11302BDD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.

Reason for recall

Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.

Key facts

Status
Ongoing
Initiation date
2020-07-27
Report date
2020-09-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
El Segundo, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2895-2020