Duo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107 AAC02108
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Nationwide distribution; worldwide distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2932-2018