# FDA recall Z-2953-2024

> **PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.** · Class II · device recall initiated 2024-08-06.

## Product

Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224

## Reason for recall

The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.

## Distribution

Worldwide - US Nationwide distribution.

## Key facts

- **Recall number:** Z-2953-2024
- **Recalling firm:** PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-08-06
- **Report date:** 2024-09-11

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Best, Netherlands