FDA recall Z-2954-2018

GE Medical Systems China Co., Ltd. · Class II · device

Product

GE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems

Reason for recall

On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.

Distribution

Worldwide

Key facts

Status
Terminated
Initiation date
2018-07-24
Report date
2018-09-05
Termination date
2020-12-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wuxi, N/A, China

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2954-2018