FDA recall Z-2961-2018

Bayer Medical Care, Inc. · Class II · device

Product

Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288

Reason for recall

The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

Distribution

The devices were distributed to the following US states: OH, TX, and WV.

Key facts

Status
Terminated
Initiation date
2018-06-28
Report date
2018-09-12
Termination date
2018-11-26
Voluntary/Mandated
Voluntary: Firm initiated
Location
Indianola, PA, United States