Zenith Alpha Abdominal Endovascular Graft - Distributed OUS only. Indicated for the endovascular treatment of patients with abdominal aortic or aorto-iliac aneurysms Reference Part Number (RPN): ZIMB-22-108 ZIMB-22-118 ZIMB-22-128 ZIMB-22-70 ZIMB-22-84 ZIMB-22-98 ZIMB-24-108 ZIMB-24-118 ZIMB-24-128 ZIMB-24-70 ZIMB-24-84 ZIMB-24-98 ZIMB-26-108 ZIMB-26-118 ZIMB-26-128 ZIMB-26-70 ZIMB-26-84 ZIMB-26-98 ZIMB-28-108 ZIMB-28-118 ZIMB-28-128 ZIMB-28-70 ZIMB-28-84 ZIMB-28-98 ZIMB-30-108 ZIMB-30-118 ZIMB-30-128 ZIMB-30-70 ZIMB-30-84 ZIMB-30-98 ZIMB-32-108 ZIMB-32-108 ZIMB-32-118 ZIMB-32-128 ZIMB-32-70 ZIMB-32-84 ZIMB-32-98 ZIMB-36-108 ZIMB-36-118 ZIMB-36-128 ZIMB-36-84 ZIMB-36-98
May contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU
Foreign Only: Canada SWEDEN AUSTRIA Belgium CZECH REPUBLIC DENMARK ESPA¿A FINLAND France GERMANY GREECE HUNGARY IRELAND ISRAEL Italia Nederland NORWAY POLAND PORTUGAL REUNION SERBIA SLOVENIA SOUTH AFRICA SWEDEN UNITED ARAB EMIRATES UNITED KINGDOM AUSTRALIA Hong Kong JAPAN NEW CALEDONIA New Zealand Thailand