FDA recall Z-2964-2020

Cook Inc. · Class II · device

Product

Zenith Alpha Abdominal Endovascular Graft - Distributed OUS only. Indicated for the endovascular treatment of patients with abdominal aortic or aorto-iliac aneurysms Reference Part Number (RPN): ZIMB-22-108 ZIMB-22-118 ZIMB-22-128 ZIMB-22-70 ZIMB-22-84 ZIMB-22-98 ZIMB-24-108 ZIMB-24-118 ZIMB-24-128 ZIMB-24-70 ZIMB-24-84 ZIMB-24-98 ZIMB-26-108 ZIMB-26-118 ZIMB-26-128 ZIMB-26-70 ZIMB-26-84 ZIMB-26-98 ZIMB-28-108 ZIMB-28-118 ZIMB-28-128 ZIMB-28-70 ZIMB-28-84 ZIMB-28-98 ZIMB-30-108 ZIMB-30-118 ZIMB-30-128 ZIMB-30-70 ZIMB-30-84 ZIMB-30-98 ZIMB-32-108 ZIMB-32-108 ZIMB-32-118 ZIMB-32-128 ZIMB-32-70 ZIMB-32-84 ZIMB-32-98 ZIMB-36-108 ZIMB-36-118 ZIMB-36-128 ZIMB-36-84 ZIMB-36-98

Reason for recall

May contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU

Distribution

Foreign Only: Canada SWEDEN AUSTRIA Belgium CZECH REPUBLIC DENMARK ESPA¿A FINLAND France GERMANY GREECE HUNGARY IRELAND ISRAEL Italia Nederland NORWAY POLAND PORTUGAL REUNION SERBIA SLOVENIA SOUTH AFRICA SWEDEN UNITED ARAB EMIRATES UNITED KINGDOM AUSTRALIA Hong Kong JAPAN NEW CALEDONIA New Zealand Thailand

Key facts

Status
Terminated
Initiation date
2020-08-11
Report date
2020-09-23
Termination date
2021-07-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bloomington, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2964-2020