FDA recall Z-2967-2020

Tosoh Bioscience Inc · Class II · device

Product

Tosoh Bioscience AIA- PACK HCG Calibrator Set-In Vitro Diagnostic for BHCG Part number 020361 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay

Reason for recall

Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.

Distribution

Worldwide Distribution: US (Nationwide) and countries of: Bogota, Cayman Islands, Mexico, Peru, and San Salvador.

Key facts

Status
Terminated
Initiation date
2020-08-07
Report date
2020-09-23
Termination date
2021-06-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Grove City, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2967-2020