FDA recall Z-2987-2018

GE Medical Systems Ultrasound & Primary Care Diagnostics, LL · Class II · device

Product

GE Vivid E90 ultrasound system Product Usage: Vivid E80 / E90 / E95 ultrasound system is general-purpose ultrasound systems, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of fetal/Obstetrics; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate), Transesophageal; Transrectal (TR); Transvaginal (TV); and lntraoperative (abdominal, thoracic, & vascular).

Reason for recall

The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.

Distribution

Nationwide and Puerto Rico FOREIGN: Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State of Brazil, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Congo, The Democratic Republic of the Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, The Former Yugoslav Republic of, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia South Africa, Spain

Key facts

Status
Terminated
Initiation date
2018-03-22
Report date
2018-09-12
Termination date
2021-08-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Milwaukee, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2987-2018