FDA recall Z-3005-2024

MEDLINE INDUSTRIES, LP - Northfield · Class II · device

Product

Medline Convenience kits labeled as: 1) URETEROSCOPY CDS SC-LF, Pack Number CDS983922F ; 2) URETEROSCOPY TRAY-RF, Pack Number DYNJ27442U ; 3) URETHROPLASTY PACK, Pack Number DYNJ39700B ; 4) RRP PACK, Pack Number DYNJ44109B ; 5) PROSTATECTOMY PACK-LF , Pack Number DYNJ66109A ; 6) PERINEAL PACK , Pack Number DYNJ69895; 7) LIVER 2018, Pack Number DYNJ905655A; 8) UROLOGY PACK, Pack Number DYNJT2255M

Reason for recall

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2024-04-08
Report date
2024-09-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3005-2024