FDA recall Z-3012-2020

Irvine Biomedical Inc, a St. Jude Medical Co. · Class II · device

Product

Inquiry Steerable Diagnostic Catheter, REF: IBI-87002 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.

Reason for recall

Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay

Distribution

Worldwide distribution - U.S. Nationwide distribution including the states of MO, IN, WI, TX, AZ, MI, VA, KY and the country of Japan.

Key facts

Status
Terminated
Initiation date
2020-08-24
Report date
2020-09-30
Termination date
2023-07-31
Voluntary/Mandated
Voluntary: Firm initiated