FDA recall Z-3031-2020

Ameditech Inc · Class II · device

Product

Alere iCup Drug Detection (AMT Item Number/ AMT Description): I-DXA-10/ "Alere" iCup¿ Dx 10 Drug Detection + Adulteration AMP/BAR/BZO/COC/MET/MDMA/MTD/MOP/OXY/THC Bl, Cr, Ni, pH, SG", I-DXA-13B/"Alere" iCup¿ DX Drug Screen Cup 13 Drug Detection + Adulteration AMP/BAR/BZO/BUPG/COC/MET/MDMA/MTD/MOP/OXY/PCP/THC/TCA Bl, Cr, Ni, pH, SG", I-DXA-1127-023-N "Alere" iCup¿ Dx Drugs of Abuse Test (COC/THC)+(MET1000/OPI2000)+(MTD/TCA)+ (OXY/MDMA)+(PCP/AMP)+BAR+BZO With Cr, Ni, pH, Bl, S.G."

Reason for recall

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Distribution

US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain

Key facts

Status
Terminated
Initiation date
2020-08-18
Report date
2020-10-07
Termination date
2024-09-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3031-2020